C-Link Protocol Commands
Measurements
Thermo Fisher Scientific
Model 48
i
Instruction Manual
B-11
Selection
Averaging Time (seconds)
7 90
8 120
9 180
10 240
11 300
co
high co
low co
These commands report the measured CO concentration when operating
in single range, or high and low CO when operating in dual or auto range
mode. The following example shows that the CO concentration is 40 ppm.
Send:
CO
Receive: CO
4.000E+1
ppm
o2
This command reports the oxygen concentration in percent. The following
example reports that the current O
2
concentration is 15%.
Send:
o2
Receive:
o2 15.00 %
o2 corr conc
This command reports the oxygen correction concentration in percent.
The following example shows that the O
2
correction concentration is 15%.
Send:
o2 corr conc
Receive:
o2 corr conc 15.00 %
set o2 corr conc
value
value
= 0 to 20.9
This command sets the oxygen correction concentration value to
value
,
where
value
is a floating-point number representing the oxygen correction
concentration in percent. The following example sets the O
2
corrected
concentration value to 15.00%.
Send:
set o2 corr conc 15
Receive:
set o2 corr conc 15 ok
flow
This command reports the current sample flow. The following example
reports that the current sample flow is 1.108 liters/minute.
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......