Operation
Service Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-85
●
In the Main Menu, choose Service >
Preamp Calibration
.
PREAMP BOARD CAL:
SAMPLE: 127807 Hz
REFERENCE: 148256 Hz
SET TO: 128 ?
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Temperature Calibration screen allows the user to view and set the
ambient temperature sensor offset, which basically adjusts the temperature
sensor so that it reads correctly. The temperature calibration is visible only
when the instrument is in service mode. For more information on the
service mode, see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
Note
Wait at least 30 seconds for the reading to stabilize before saving the
value.
▲
●
In the Main Menu, choose Service >
Temperature Calibration
.
CALIBRATE AMBIENT TEMP:
CURRENTLY: 32.3
o
C
SET TO: 032.
5
o
C ?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Analog Output Calibration menu provides access to the 6 voltage
channels and 6 current channels for calibration. Current channels are
visible only if the I/O expansion board is installed. The analog output
calibration menu is visible only when the instrument is in service mode.
For more information on the service mode, see “
Service Mode
” earlier in
the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
Temperature
Calibration
Analog Output
Calibration
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......