C-Link Protocol Commands
Keys/Display
Thermo Fisher Scientific
Model 48
i
Instruction Manual
B-33
zs avg time
This command reports the zero/span (zs) averaging time in seconds. The
following example reports that the zero/span averaging time is 30 seconds,
according to
Table B–3
.
Send:
zs
avg
time
Receive:
zs avg time 5:30 sec
set avg time
value
This command sets the zero/span averaging time, according to
Table B–3
.
The following example sets the zero/span averaging time to 120 seconds.
Send:
set zs avg time 8
Receive:
set zs avg time 8 ok
push
button
do
ri
down
right
en
ru
enter
run
he
up
help
1
le
2
left
3
me
4
menu
button
= | do | down | en | enter | he | help | le | left | me | menu | ri | right |
ru | run | up | 1 | 2 | 3 | 4 |
These commands simulate pressing the front panel pushbuttons. The
numbers represent the front panel soft keys, from left to right.
Send:
push
enter
Receive:
push enter ok
isc
iscreen
These commands retrieve the framebuffer data used for the display on the
i
Series instrument. It is 19200 bytes in size, 2-bits per pixel, 4 pixels per
byte arranged as 320 by 240 characters. The data is sent in RLE encoded
form to save time in transmission. It is sent as a type '5' binary C-Link
response with no checksum.
The RLE encoding consists of a 0 followed by an 8-bit count of
consecutive 0xFF bytes. The following 'c' code will expand the incoming
data.
Keys/Display
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......