Thermo Fisher Scientific
Model 48
i
Instruction Manual
D-1
Appendix D
Gesytec (Bayern-Hessen) Protocol
This appendix provides a description of the Gesytec (Bayern-Hessen or
BH) Protocol Interface and is supported both over RS-232/485 as well as
TCP/IP over Ethernet.
The Gesytec commands that are implemented are explained in detail in this
document. The Gesytec protocol support for the
i
Series enables the user to
perform the functions of reading the various concentrations and to trigger
the instrument to be in sample/zero/span mode if valid for that instrument.
This is achieved by using the supported Gesytec commands listed below.
For details of the Model 48
i
Gesytec Protocol specification, see the
following topics:
●
“
Serial Communication Parameters
” on page
D-1
●
“
TCP Communication Parameters
” on page
D-2
●
“
Instrument Address
” on page
D-2
●
"
Abbreviations Used
” on page
D-2
●
“
Basic Command Structure
” on page
D-2
●
“
Block Checksum Characters <BCC>
” on page
D-3
●
“
Gesytec Commands
” on page
D-3
The following are the communication parameters that are used to configure
the serial port of the
i
Series to support Gesytec protocol.
Number of Data bits
: 7 or 8
Number of Stop bits
: 1 or 2
Parity
: None, Odd, or Even
Data rate
: 1200 to 115200 Baud (9600 is default)
Serial Communication
Parameters
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......