Operation
Alarms Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-73
The AGC Intensity screen displays the current AGC Intensity reading and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 150,000 to 300,000 Hz. If the AGC intensity reading goes
beyond either the minimum or maximum limit, an alarm is activated and
the alarm (bell) icon appears in the status bar on the Run screen and in the
Main Menu.
●
In the Main Menu, choose Alarms >
AGC Intensity
.
AGC INTENSITY:
ACTUAL 196646 Hz
>MIN 150000 Hz
MAX 300000 Hz
RANGE AVG DIAGS ALARM
The Minimum AGC Intensity alarm limit screen is used to change the
minimum AGC intensity alarm limit. The minimum and maximum AGC
Intensity screens function the same way.
●
In the Main Menu, choose Alarms > AGC Intensity >
Min
or
Max
.
AGC INTENSITY:
ACTUAL MIN: 150000 Hz
SET MIN TO: 200000 Hz ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Motor Speed screen (read only) displays the current motor speed. A
reading other than 100.0% indicates a problem with either the motor or
the power supply.
●
In the Main Menu, choose Alarms >
Motor Speed
.
AGC Intensity
Min and Max AGC
Intensity Limits
Motor Speed
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......