C-Link Protocol Commands
Measurement Configuration
B-36
Model 48
i
Instruction Manual
Thermo Fisher Scientific
custom
range
range
= | 1 | 2 | 3 |
This command reports the user-defined value of custom range 1, 2, or 3.
The following example reports that custom range 1 is defined to 5.50 ppm.
Send:
custom
1
Receive:
custom 1 5.500E+3 ppm
set custom
range
range
value
set custom 1 range
value
set custom 2 range
value
set custom 3 range
value
These commands are used to set the maximum concentration for any of the
three custom
ranges
1, 2, or 3 to range
value
, where
value
is a floating-point
number representing concentration in ppm or mg/m
3
. The following
example sets the custom 1 range to 55.5 ppm.
Send:
set custom 1 range 55.5
Receive:
set custom 1 range 55.5 ok
range mode
This command reports the current range mode.
Send:
range
mode
Receive:
range mode single
set range mode
mode
This command sets the current range mode to single, dual, or auto. The
following example sets the range mode to single.
Send:
set range mode single
Receive:
set range mode single ok
gas mode
This command reports the current state of the zero/span valves gas mode as
sample, zero, or span. The following example reports that the gas mode is
sample.
Send:
gas
mode
Receive: gas
mode
sample
set sample
This command sets the zero/span valves to the sample mode. The following
example sets the instrument to sample mode, that is, the instrument is
reading the sample gas.
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......