Operation
Alarms Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-75
●
In the Main Menu, choose Alarms >
Zero
or
Span Autocal
.
ZERO AUTO CALIBRATION:
ALARM: OK
RESPONSE: 0.200
RANGE AVG DIAGS ALARM
The Concentration screen displays the current CO concentration and sets
the minimum and maximum alarm limits. Acceptable alarm limits range
from 0 to 10000. The minimum alarm may be programmed as a floor
trigger (alarm is triggered when the concentration falls below the minimum
value) or a ceiling trigger (alarm is triggered when the concentration goes
above the minimum value). If the CO concentration goes beyond either
the minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Concentration
.
CONCENTRATION:
ACTUAL 0.000
>MIN 10000
MAX 10000
MIN TRIGGER CEILING
RANGE AVG DIAGS ALARM
The Minimum Concentration alarm limit screen is used to change the
minimum concentration alarm limits. The minimum and maximum
concentration alarm limit screens function the same way.
●
In the Main Menu, choose Alarms > Concentration >
Min
or
Max
.
CONCENTRATION:
ACTUAL MIN: 10000
SET MIN TO: 0100.00
0
?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Concentration
Min and Max
Concentration Limits
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......