Operation
Calibration Menu
3-20
Model 48
i
Instruction Manual
Thermo Fisher Scientific
●
In the Main Menu, choose Calibration Factors >
O
2
Coefficient
.
O2 COEFFICIENT:
O2: 20.80
SET COEF TO: 1.00
?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Reset User Calibration Defaults screen allows the user to reset the
calibration configuration values to factory defaults.
●
In the Main Menu, Calibration Factors >
Reset User Cal Defaults
.
The Calibration menu is used to calibrate zero and span. The Calibrate O
2
Background menu item and the Calibrate O
2
Coefficient menu item are
displayed only when the internal O
2
sensor option is installed.The
calibration menu is similar for the single, dual, and auto range mode as
shown below. The only difference between the screens are the words “HI”
and “LO” to indicate which range is displayed. The dual and auto range
modes have two CO span factors (high and low). This allows each range to
be calibrated separately. This is necessary if the two ranges used are not
close to one another. For example, a low CO range of 50 ppm and a high
CO range of 1,000 ppm. For more information about calibration, see
Chapter 4, “
Calibration
”.
●
In the Main Menu, choose
Calibration
.
Reset User
Calibration Defaults
Calibration
Menu
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......