C-Link Protocol Commands
Communications Configuration
Thermo Fisher Scientific
Model 48
i
Instruction Manual
B-47
Table B–11.
Power Up Mode Values
Value
Power up Mode
0 Local/Unlocked
Mode
(default)
1 Remote/Locked
Mode
program no
This command reports the analyzer’s model information and program
version number.
Send:
program
no
Receive:
program no iSeries 48i 01.02.12.095
set layout ack
This command reports the stale layout/layout change indicator (*) that is
attached to each response if the erec layout has changed since the last time
erec layout was requested, according to
Table B–12
. The following example
reports that the instrument is configured to do nothing.
Send:
layout
ack
Receive: layout
ack
0
set layout ack
value
This command disables the stale layout/layout change indicator (*) that is
attached to each response if the erec layout has changed since the last time
erec layout was requested, according to
Table B–12
.
Send:
set
layout
ack
Receive:
set layout ack ok
Table B–12.
Set Layout Ack Values
Value
Function
0
Do nothing (default)
1 Append
“*”
tz
This command reports the “tz” timezone string for the NTP server. See
“Network Time Protocol Server” in the “Communications Settings”
section of the “Operation” chapter for more information.
Send:
tz
Receive: tz
EST+5EDT
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part Number 101891 00 10Aug2017 ...
Страница 4: ......
Страница 6: ......
Страница 37: ...Installation Connecting External Devices Thermo Fisher Scientific Model 48i Instruction Manual 2 9 ...
Страница 144: ......
Страница 154: ......
Страница 216: ......
Страница 228: ......
Страница 237: ...Optional Equipment Mounting Options Thermo Fisher Scientific Model 48i Instruction Manual 9 9 Figure 9 4 Bench Mounting ...
Страница 240: ......
Страница 321: ......