Servicing
LCD Module Replacement
7-40
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Figure 7–18.
Replacing the Front Panel Board and the LCD Module
Use the following procedure to replace the LCD module (
Figure 7–18
).
Equipment Required:
LCD module
Philips screwdriver
CAUTION
If the LCD panel breaks, do not let the liquid crystal contact
your skin or clothes. If the liquid crystal contacts your skin or clothes, wash
immediately using soap and water.
▲
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
Do not remove the LCD panel or frame from the LCD module.
▲
The LCD polarizing plate is very fragile, handle it carefully.
▲
Do not wipe the LCD polarizing plate with a dry cloth, as it may easily
scratch the plate.
▲
LCD Module
Replacement
LCD Module
LCD Module Screws (2)
Front Panel Board
Mounting Studs (2)
Ribbon Cables (3)
Bottom Support
Two-wire Connector
Retaining Screws (2)