Service Maintenance
Replacing the Turbomolecular Pump _____________________________________________
5-64
___________________ LCQ Deca XP Plus Hardware Manual ________________
Reinstalling the Turbomolecular Pump
To reinstall the turbomolecular pump, proceed as follows. See Figure 5-18.
1. Check the condition of the Viton O-rings around the two openings on the
bottom of the vacuum manifold. (Use a small flashlight to illuminate the
O-rings.) If they have any nicks or breaks, replace them with new ones
(P/N 00107-11100).
2. Place the turbomolecular pump on the turbomolecular pump rails.
3. Slide the turbomolecular pump into position under the openings in the
vacuum manifold.
4. With a 3/16-in. ball driver or Allen wrench, carefully tighten the two
Allen screws that hold the rails and turbomolecular pump to the vacuum
manifold. Do not overtighten the screws.
5. Reconnect the vacuum hose to the tee (P/N 97144-20031). Tighten the
hose clamp that secures the vacuum hose to the tee.
6. Reconnect the turbomolecular pump power cable to the turbomolecular
pump controller.
7. Reinstall AT box.
8. If necessary, reconnect the vacuum hoses to the foreline union. Tighten
the clamping rings to secure the vacuum hoses to the foreline union.
9. Reinstall the top cover of the MS detector by following the procedure in
the topic Reinstalling the Top Cover of the MS Detector on
page 5-16.
10. Close the left and right front doors of MS detector.
11. Reconnect any tubing between the syringe pump and the API source that
you disconnected earlier.
12. Restart the system as described in the topic Starting Up the System after
a Complete Shutdown in the System Shutdown, Startup, and Reset
chapter.