Service Maintenance
________________________________________ Cleaning the Ion Optics and Mass Analyzer
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-21
c.
Allow the program to make at least five passes. Then determine
whether the frequency cursor lies within the frequency window:
•
If the frequency cursor lies within the frequency window, then the
ring electrode RF voltage frequency is tuned properly. Click on
Stop
and exit from the diagnostics program.
•
If the frequency cursor lies outside the frequency window, then
you need to manually adjust the ring electrode RF voltage
frequency. Leave the Graph view displayed. Go on to the next
step.
Figure 5-7.
Graph view for ring electrode RF voltage tuning
Frequency
Cursor
Frequency
Window