User Maintenance
______________________________________________________ API Source Maintenance
__________________ LCQ Deca XP Plus Hardware Manual _________________
4-13
Removing the APCI Probe Assembly
If the APCI probe assembly is installed on the XP Plus, first remove the APCI
probe assembly as described in the topic Removing the APCI Probe
Assembly in the Changing API Probe Assemblies chapter. Place the APCI
probe assembly on a lint-free tissue.
Go on to the next topic: Disassembling the APCI Probe Assembly.
Disassembling the APCI Probe Assembly
To disassemble the APCI probe assembly, do the following:
1. Remove the corona discharge needle (P/N 70005-98033) by pulling it free
from the corona discharge needle assembly. Store the needle in a safe
place.
2. Remove the APCI probe from the APCI flange as follows:
a.
Unbolt the APCI slider assembly from the APCI flange. You will not
be able to remove the probe from the flange at this time because the
cable from the APCI vaporizer heater is still connected to the flange.
CAUTION. AVOID BURNS. The APCI vaporizer heater can reach
temperatures of 600
°
C. Always allow the APCI probe to cool to ambient
temperatures before handling or removing the APCI probe from the APCI
flange.
Caution. Wrench flats on the APCI probe components are provided for
your convenience when you dismantle the APCI probe. NEVER USE THE
WRENCH FLATS TO TIGHTEN THE APCI PROBE
COMPONENTS. Only tighten the APCI probe components by hand.
Caution. Do not break the APCI sample tube. In step 2a, carefully pull
the APCI manifold straight back from the APCI probe to prevent the sample
tube from touching the sides. If the sample tube hits the sides of the sheath
gas tube, it can break.