Service Maintenance
____________________________________________________________ Replacing a Fuse
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-33
3. To replace a fuse on the System Control PCB, proceed as follows. See
Figure 5-11 for the location of the System Control PCB and its fuses.
a.
Remove the right side cover of the MS detector, as follows:
i.
Loosen the fastener that secures the right side cover to the chassis
of the MS detector.
ii. Slide the side cover back about 1.25 cm (0.5 in.), and then pull it
out and away from the MS detector.
b. Remove the System Control PCB protective cover, as follows:
i.
With a Phillips screwdriver, loosen the five screws that hold the
protective cover to the MS detector chassis.
ii. Pull the cover up and out of the guide slots on the MS detector
chassis.
c.
Locate and replace the defective fuse on the System Control PCB
with a fuse of the same type, voltage, and current rating. Refer to
Table 5-5.
Caution. To prevent damage to the electronics due to electrostatic
discharge, attach an electrostatic discharge (ESD) strap to your wrist before
continuing.
Figure 5-12. System Control PCB, showing the location of the fuses
Fuses