User Maintenance
__________________________________________________________________________
4-2
____________________ LCQ Deca XP Plus Hardware Manual ________________
For instructions on maintaining LCs or autosamplers, refer to the manual that
comes with the LC or autosampler.
The topics included in this chapter are as follows:
•
Tools and supplies
•
Frequency of cleaning
•
API source maintenance
Note. The keys to success with the procedures in this chapter are:
•
Proceed methodically.
•
Always wear clean, lint-free gloves when handling the components of the
API source, ion optics, mass analyzer, and ion detection system.
•
Always place the components on a clean, lint-free surface.
•
Always cover the opening in the top of the vacuum manifold with a large,
lint-free tissue whenever you remove the top cover plate of the vacuum
manifold.
•
Never overtighten a screw or use excessive force.
•
Never insert a test probe (for example, an oscilloscope probe) into the
sockets of female cable connectors on PCBs.