User Maintenance
API Source Maintenance ______________________________________________________
4-12
___________________ LCQ Deca XP Plus Hardware Manual ________________
Figure 4-1 shows the major components of the APCI probe. You do not need
to vent the system to perform maintenance on the APCI probe.
The following procedures are discussed in this section:
•
Removing the APCI probe assembly
•
Disassembling the APCI probe assembly
•
Cleaning the APCI probe components
•
Removing the APCI sample tube
•
Installing the APCI sample tube
•
Reassembling the APCI probe assembly
•
Reinstalling the APCI probe assembly
Note. You should flush the APCI probe at the end of each working day by
flowing a 50:50 methanol/water solution from the LC through the APCI
source. Refer to the topic Flushing the Sample Transfer Line, Sample
Tube, and API Probe on page 4-6.
Note. Wear clean gloves when you handle APCI probe components.
Figure 4-1.
Cross sectional view of the APCI probe
Probe
Collar
APCI
Nozzle
Ground
Wire
Adjustment
Handle
Vaporizer
Casing
Quartz
Insulator
Heater
Coil
Vaporizer
Tube
APCI
Manifold
Sheath
Gas Inlet
Heater
Retainer
Auxiliary
Gas Inlet
Sample Tube
Inlet Fitting
Sample
Tube