![Thermo Finnigan LCQ Deca XP Plus Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/thermo-finnigan/lcq-deca-xp-plus/lcq-deca-xp-plus_hardware-manual_1104316030.webp)
Functional Description
API Source _________________________________________________________________
1-6
____________________ LCQ Deca XP Plus Hardware Manual ________________
Figure 1-3.
Geometry of the ESI probe and API stack
Ion
Transfer
Capillary
Heater
To Ion
Skimmer
Pressure
= 1 Torr
+10 V
Atmospheric
Pressure
Region
Ion
Sweep
Cone
+ + +
+
+
+
+
+ +
+
+
+
–
–
–
–
–
–
+ + +
+
+
+
–
+
+
+ +
–
– – –
––
–
Charged
Droplet
Solvent
Evaporates
From
Droplet
Ejected
Positive
Ions
ESI Needle
(+5 kV)
ESI Nozzle