Service Maintenance
___________________________________________ Replacing the RF Voltage Control PCB
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-57
5.9 Replacing the RF Voltage Control
PCB
To replace the RF Voltage Control PCB (and its housing), proceed as follows.
See Figure 5-16 for the location of the RF Voltage Control PCB.
1. Place the electronics service switch in the Service Position (or shut down
and vent the XP Plus system as described in the topic Shutting Down the
System Completely in the System Shutdown, Startup, and Reset
chapter).
Figure 5-16. Left side of the MS detector, showing the RF Voltage
Control PCB
RF Voltage Control
PCB