Service Maintenance
___________________________________________ Replacing the RF Voltage Control PCB
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-59
12. Reconnect the cable that comes from the RF Voltage Amplifier PCB.
13. Position the front cover over the RF Voltage Control PCB. With a
Phillips screwdriver, reinstall the screws that hold the front cover to the
RF Voltage Control PCB housing.
14. Reinstall the left side cover of the MS detector as follows:
a.
Place the cover against the left side of the MS detector such that the
studs on the cover insert into the guide slots in the MS detector
chassis.
b. Slide the side cover forward about 1.25 cm (0.5 in.) until the studs on
the cover lock in the guide slots.
c.
Tighten by hand the fastener that secures the side cover to the chassis
of the MS detector.
15. Reinstall the top cover of the MS detector as described in the topic
Reinstalling the Top Cover of the MS Detector on page 5-16.
16. Place the electronics service switch in the Operating Position.
17. Run the Tune Plus diagnostics to verify that the system is operational.
Replacing the Low Pass Filter PCB
To replace the Low Pass Filter PCB, proceed as follows. See Figure 5-17 for
the location of the Low Pass Filter PCB.
1. Place the electronics service switch in the Service Position (or shut down
and vent the XP Plus system as described in the topic Shutting Down the
System Completely in the System Shutdown, Startup, and Reset
chapter).
2. Remove the top cover of the MS detector as described in the topic
Removing the Top Cover of the MS Detector on page 5-6.
3. Remove the rear cover of the MS detector as follows:
a.
With a Phillips screwdriver, loosen the 10 screws that secure the rear
cover to the chassis of the MS detector.
b. Slide the rear cover up about 1.25 cm (0.5 in.), and then lift it out and
away from the MS detector.
CAUTION. Make sure that the XP Plus electronic service switch is in the
Service Position (or shut down or shut down the system and disconnect the
power cord) before you proceed.