Service Maintenance
Cleaning the Ion Optics and Mass Analyzer _______________________________________
5-16
___________________ LCQ Deca XP Plus Hardware Manual ________________
7. Reconnect (at ANAL. AUX 2 IN and ANAL. AUX 3 IN) the two endcap
electrode cables that come from the Analyzer Auxiliary PCB.
8. Reconnect (at ACQU/DSP) the electrometer cable.
9. Reconnect (at MULT) the electron multiplier high voltage cable that
comes from the electron multiplier power supply.
Reinstalling the Top Cover of the MS
Detector
Reinstall the top cover of the MS detector, as follows:
1. Open the left and right front doors of the MS detector.
2. With one hand under the center of the top cover, place the top cover on the
MS detector chassis. Slide the top cover forward until it is engages the
four guides located at the rear of the chassis.
3. Lower the front of the top cover onto the chassis so that the two guide
posts located on the front underside of the top cover enter the guide holes
located on the top of the chassis.
4. Secure the top cover to the chassis with the two latches located on the
front of the chassis.
5. Close the left and right front doors of the MS detector.
6. Reconnect any tubing between the syringe pump and the API source to
accommodate your instrument configuration.
Starting Up the System
Start up the system as described in the topic Starting Up the System after a
Complete Shutdown in the System Shutdown, Startup, and Reset chapter.
Tuning the Ring Electrode and
Quadrupole/Octapole RF Voltages
You need to tune the ring electrode RF voltage and the quadrupole/octapole
RF voltage whenever you service the mass analyzer or ion optics. You also
need to tune these voltages if you replace any electronic assembly that is
involved in producing the RF voltages. You use the Diagnostics program to
tune the ring electrode and quadrupole/octapole RF voltages.