Service Maintenance
Replacing PCBs and Assemblies________________________________________________
5-36
___________________ LCQ Deca XP Plus Hardware Manual ________________
5.7 Replacing PCBs and Assemblies
The following procedures are discussed in this topic:
•
Replacing PCBs and assemblies in the tower
•
Replacing PCBs in the embedded computer
•
Replacing the Vent Delay PCB and backup battery, ion gauge, and vent
valve
•
Replacing the electron multiplier and Conversion Dynode Power
Supplies, Analyzer PCB, Analyzer Auxiliary PCB, Waveform Amplifier
PCB, RF Voltage Amplifier PCB
•
Replacing the System Control PCB
•
Replacing the RF Voltage Control PCB
•
Replacing the Low Pass Filter PCB
CAUTION. With the electronics service switch in the Service Position, the
system is still under vacuum. Power is still supplied to the rotary-vane
pump, Turbomolecular Pump Controller, turbomolecular pump,
turbomolecular pump fan, +24 V keep alive power supply, vent valve, and
Vent Delay PCB. Thus, before these components can be serviced, the main
power circuit breaker switch must be placed in the Off (O) position and the
power cord must be unplugged from the power outlet.
Caution. Never insert a test probe (for example, an oscilloscope probe) into
the sockets of female cable connectors on PCBs. This can damage the
sockets.