User Maintenance
______________________________________________________ API Source Maintenance
__________________ LCQ Deca XP Plus Hardware Manual _________________
4-35
9. Reconnect the skimmer lead to the lead pin on the skimmer. Use
needlenose pliers if necessary.
10. Inspect the API stack. Ensure that the 3.85-in. ID O-ring
(P/N 00107-14100) has no cracks, is otherwise in good condition, and is
properly seated on the spray shield. Ensure that all components fit
together tightly.
11. Reinstall the ion transfer capillary. See Figure 4-10. Replace the Kalrez
O-ring (P/N 00107-12750) if necessary. Refer to the topic Maintaining
the API Stack on page 4-27.
12. Reinstall the ion sweep cone (if being used).
Go on to the next topic: Reinstalling the API Stack.
Figure 4-10. Exploded view of the front of the API stack
Ion Sweep
Cone
Ion Transfer
Capillary
Kalrez
O-Ring
Spray Cone