Changing API Probe Assemblies
Removing the APCI Probe Assembly_____________________________________________
2-2
____________________ LCQ Deca XP Plus Hardware Manual ________________
2.1 Removing the APCI Probe
Assembly
To remove the APCI probe assembly, proceed as follows. Refer to
Figure 2-1 for the location of the components of the APCI probe assembly.
1. Stop the flow of sample solution (from the LC or syringe pump) into the
APCI probe.
2. Place the electronics service switch in the Service position. When you
place the electronics service switch in the Service position, the voltages to
the API source (including the APCI probe), ion optics, mass analyzer, and
ion detection system are turned off and the flows of sheath gas and
AUX/Sweep gas (if used) into the APCI probe are stopped.
3. Allow APCI probe and flange to cool to ambient temperature before
proceeding.
4. Loosen the two flange retainer bolts that secure the APCI probe assembly
to the API spray shield. See Figure 2-1.
5. Pull the APCI probe assembly (still attached to the probe guide) back
from the spray shield.
6. Disconnect the corona needle high voltage cable from the corona needle
high voltage connector. To disconnect the cable, rotate the locking ring on
the cable counterclockwise until you can pull the cable free.
7. Disconnect the vaporizer heater cable from the connector on the front
fascia panel. (Leave the cable connected to the APCI flange). To
disconnect the cable, rotate the locking ring on the cable
counterclockwise until you can pull the cable free.
8. Disconnect the sample transfer line from the APCI probe by turning the
sample transfer line fitting counterclockwise until you can pull the
transfer line and fitting free from the probe.
CAUTION. Make sure that the LCQ Deca XP Plus electronics service
switch is in the Service position before you proceed.
CAUTION. The APCI vaporizer heater can reach temperatures of 600
°
C.
Always allow the APCI probe and flange to cool to ambient temperatures
before handling or removing the APCI probe or flange.