Service Maintenance
________________________________________ Cleaning the Ion Optics and Mass Analyzer
__________________ LCQ Deca XP Plus Hardware Manual __________________
5-9
1. Prepare a clean work area by covering the area with lint-free paper. Place
each part on the paper as you remove it.
2. Disconnect the four electrical leads to the quadrupole.
3. Hold the quadrupole mount with one hand; loosen and remove the two
thumb screws that hold the quadrupole mount to the baffle on the top
cover plate of the vacuum manifold.
4. Remove the quadrupole and quadrupole mount.
5. Disconnect the electrical lead to the interoctapole lens. Remove the
interoctapole lens.
6. Disconnect the electrical leads to the octapole and to the entrance lens,
entrance endcap electrode, exit endcap electrode, and the exit lens of the
mass analyzer.
7. Disconnect the damping gas line from the nipple on the exit endcap
electrode by pulling the line free from the nipple. See Figure 5-2.
8. Hold the mass analyzer with one hand; loosen the two thumb screws that
hold the analyzer mount to the baffle.
9. With one hand holding the mass analyzer and the other hand holding the
analyzer mount, lift the mass analyzer, octapole, and analyzer mount out
and away from the baffle on the top cover plate. Be careful not to touch
the electron multiplier with the mass analyzer. This could damage the
electropolished surface.
Disassembling the Ion Optics and Mass
Analyzer
To disassemble the ion optics and mass analyzer, proceed as follows. See
Figure 5-2 for the location of the ion optics and mass analyzer components.
1. Remove the quadrupole from the quadrupole mount.
2. Remove the octapole from the analyzer mount.
3. Disassemble the mass analyzer, as follows:
a.
Remove the entrance lens by pulling the entrance lens out of the
entrance lens sleeve.
b. Remove the entrance lens sleeve by squeezing the sleeve and pulling
it out of the recess in the entrance endcap electrode.
Note. Wear clean, lint-free, nylon or cotton gloves when you handle the ion
optics and mass analyzer components.