System Shutdown, Startup, and Reset
___________________Resetting the Tune and Calibration Parameters to their Default Values
__________________ LCQ Deca XP Plus Hardware Manual __________________
3-9
3.6 Resetting the Tune and Calibration
Parameters to their Default Values
You can reset the XP Plus tune and calibration parameters to their default
values at any time. This feature may be useful if you have manually set some
parameters that have resulted in less than optimum performance. To reset the
XP Plus tune and calibration parameters to their default values, proceed as
follows: In the Tune Plus window, choose
File | Restore Factory
Calibration
to restore the default calibration parameters, or choose
File |
Restore Factory Tune Method
to restore the default tune parameters.
Note. Make sure that the problems that you might be experiencing are not
due to improper API source settings (spray voltage, sheath and auxiliary gas
flow, ion transfer capillary temperature, etc.) before resetting the system
parameters to their default values.