Service Maintenance
Cleaning the Ion Optics and Mass Analyzer _______________________________________
5-6
____________________ LCQ Deca XP Plus Hardware Manual ________________
Removing the Top Cover of the MS Detector
Remove the top cover of the MS detector, as follows:
1. Disconnect any tubing between the syringe pump and the API source.
2. Open the left and right front doors of the MS detector by loosening the
1/4-in. Allen screw on the right front door with an Allen wrench.
3. Open the two fasteners that hold the top cover to the MS detector chassis.
The fasteners are located in the upper right and left corners of the chassis.
4. With one hand under the center of the top cover, lift the top cover up and
away from the MS detector.
Removing the Top Cover Plate of the Vacuum
Manifold
You need to remove the top cover plate of the vacuum manifold to access the
ion optics, mass analyzer, and ion detection system. The top cover plate is
held in place by gravity and by the air pressure differential between the
vacuum manifold and atmospheric pressure. Six cables and one gas line are
connected to the top cover plate.
To remove the top cover plate, proceed as follows. See Figure 5-1.
1. Disconnect (at ANAL. AUX 1 IN) the octapoles cable that comes from
the Analyzer Auxiliary PCB.
2. Disconnect (at ANALYZER) the lenses cable that comes from the System
Control PCB.
3. Disconnect (at ANAL. AUX 2 IN and ANAL. AUX 3 IN) the two endcap
electrode cables that come from the Analyzer Auxiliary PCB.
4. Disconnect (at ACQU/DSP) the electrometer cable. (If necessary, use a
small screw driver to loosen the screws that secure the cable.)
5. Disconnect (at MULT) the electron multiplier high voltage cable that
comes from the electron multiplier power supply.
6. Use a 7/16-in. open-end wrench to disconnect the helium damping gas
line from the fitting.
7. Carefully lift the top cover plate straight up by its two handles. Take care
not to damage the components on the underside of the cover plate. Place
the cover plate upside down (supported on its handles) on a flat surface.
8. Cover the opening in the top of the vacuum manifold with a large,
lint-free tissue.