Service Maintenance
____________________________________________________________ Replacing a Fuse
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-35
d. Reinstall the protective cover on the Analyzer Auxiliary PCB. With a
Phillips screwdriver, tighten the six screws that hold the metal cover
to the Analyzer Auxiliary PCB.
e.
Reconnect the seven cables that connect to the top of the Analyzer
Auxiliary PCB. (Three coaxial cables come from the Analyzer PCB,
one coaxial cable comes from the Waveform DDS PCB in the
embedded computer, one coaxial cable comes from the Waveform
Amplifier PCB, one 7-lead cable comes from the System Control
PCB, and one 4-lead cable comes from the 36 V power supply.)
6. Reinstall the top cover of the MS detector as described in the topic
Reinstalling the Top Cover of the MS Detector on page 5-16.
7. Place the electronics service switch in the Operating Position.
8. Run the Tune Plus diagnostics to verify that the system is operational.