Service Maintenance
Cleaning the Ion Optics and Mass Analyzer _______________________________________
5-12
___________________ LCQ Deca XP Plus Hardware Manual ________________
6. Reinstall the exit endcap electrode (the one with the damping gas nipple)
on the analyzer posts such that the quartz spacer ring is held in place
between the ring electrode and the exit endcap electrode. Make sure that
the electrode is oriented such that the convex surface faces the spacer
ring. Also, the damping gas nipple should point toward the top cover
plate when the analyzer mount is installed on the top cover plate.
7. Inspect the mass analyzer assembly. Ensure that all the parts are aligned
properly and that they all fit together snugly.
8. Reinstall the spring washers on the analyzer posts such that the convex
side of the washer is toward the exit endcap electrode.
9. Reinstall the analyzer nuts onto the analyzer posts and tighten the nuts by
hand until they are finger tight. Do not overtighten the nuts.
10. Squeeze the exit lens sleeve and insert it into the recess in the exit endcap
electrode. See Figure 5-2 for the proper orientation of the exit lens sleeve.
11. Insert the exit lens into the exit lens sleeve such that the lead pin on the
exit lens points in the same direction as the 8-pin feedthrough when the
analyzer mount is installed on the top cover plate. Make sure that the exit
lens lead pin does not contact the nut on the end of the mass analyzer post.
Reinstalling the Ion Optics and Mass
Analyzer
Use the following procedure to reinstall the ion optics and mass analyzer
onto the top cover plate of the vacuum manifold:
1. Insert the cylindrical end of the analyzer mount (with the mass analyzer
and octapole attached) into the opening in the baffle on the top cover plate
of the vacuum manifold. Ensure that the open side of the analyzer mount
is away from the top cover plate.
2. Secure the analyzer mount to the baffle with the two thumb screws.
3. Squeeze the entrance lens sleeve and insert it into the recess in the
entrance endcap electrode. See Figure 5-2 for the proper orientation of the
exit lens sleeve.
4. Insert the entrance lens into the entrance lens sleeve such that the lead pin
on the entrance lens points in the same direction as the 8-pin feedthrough
when the analyzer mount is installed on the top cover plate.
Note. Wear clean, lint-free, nylon or cotton gloves when you handle
components of the ion optics and mass analyzer.