Service Maintenance
______________________________________ Replacing PCBs and Assemblies in the Tower
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-51
n. Use a 7/16-in. open-end wrench to reconnect the helium damping gas
line to the fitting.
o. Reconnect (at P5) the octapoles cable that comes from the Analyzer
Auxiliary PCB.
p. Reconnect (at P4) the lenses cable that comes from the System
Control PCB.
q. Reconnect (at P2 and P3) the two endcap cables that come from the
Analyzer Auxiliary PCB.
r.
Reconnect (at P1) the electrometer cable.
s.
Reconnect the electron multiplier high voltage cable that comes from
the electron multiplier power supply.
t.
Go to step 8.
5. To replace the Analyzer Auxiliary PCB, proceed as follows. See
Figure 5-15 on page 5-47 for the location of the Analyzer Auxiliary PCB.
a.
Disconnect all cables to the Analyzer Auxiliary PCB.
b. With a Phillips screwdriver, remove the metal cover from the
Analyzer Auxiliary PCB.
c.
With a Phillips screwdriver, remove the screws that hold the Analyzer
Auxiliary PCB to the top of the vacuum manifold. Remove the
Analyzer Auxiliary PCB.
d. Unpack the new Analyzer Auxiliary PCB (P/N 97000-61340). Retain
the packing materials so that you can pack and ship the defective PCB
to the Thermo Finnigan Repair Center in San Jose. Be sure to note
the apparent problem or symptoms on the enclosed forms.
e.
Install the new PCB in the place occupied by the old PCB.
f.
With a Phillips screwdriver, reinstall the screws that secure the
Analyzer Auxiliary PCB to the top of the vacuum manifold.
g. Reinstall the metal cover to the top of the Analyzer Auxiliary PCB.
h. Reconnect all cables to the Analyzer Auxiliary PCB that you
disconnected in step 5a.
i.
Go to step 8.
Caution. To prevent damage to the electronics due to electrostatic
discharge, attach an electrostatic discharge (ESD) strap to your wrist before
continuing.