![Thermo Finnigan LCQ Deca XP Plus Скачать руководство пользователя страница 153](http://html.mh-extra.com/html/thermo-finnigan/lcq-deca-xp-plus/lcq-deca-xp-plus_hardware-manual_1104316153.webp)
Service Maintenance
______________________________________ Replacing PCBs and Assemblies in the Tower
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-47
4. To replace the vent valve, proceed as follows:
a.
Disconnect the cable that comes from the Vent Delay PCB.
b. With a 7/16-in. open-end wrench, loosen the fitting to the vent valve
solenoid. Remove the vent valve.
c.
Replace the old vent valve with a new one (P/N 97000-60128)
d. With a 7/16-in. open-end wrench, tighten the fitting to the vent valve
solenoid.
Figure 5-15. PCBs and assemblies that are accessible from the top of the MS detector
Conversion Dynode and
Electron Multiplier
Combo Power Supply
Vent Delay PCB
And Battery
Backup Supply
Ion Gauge
RF Voltage
Amplifier PCB
Waveform
Amplifier PCB
Analyzer
Auxiliary PCB
Vent Valve
Convectron
Gauge
Analyzer PCB