Service Maintenance
Cleaning the Ion Optics and Mass Analyzer _______________________________________
5-22
___________________ LCQ Deca XP Plus Hardware Manual ________________
6. Manually adjust the ring electrode RF voltage frequency, as follows:
a.
Open the left front door of the MS detector.
b. With a Phillips screw driver, remove the air deflector to expose the
tuning stud.
c.
With a wrench, loosen the 9/16-in. lock nut that holds the tuning stud
in place. See Figure 5-8.
d. With a screw driver, turn the tuning stud until the frequency cursor
lies slightly to the left of the center of the frequency window. (The
cursor should shift slightly to the right when the air deflector is
reinstalled.)
e.
Tighten the 9/16-in. lock nut.
f.
Reinstall the air deflector and close the left front door of the MS
detector. Make sure that the frequency cursor is still within the
frequency window. If necessary, repeat the above steps.
g. Click on
Stop
to stop the ring electrode RF voltage frequency tune
program.
Figure 5-8.
Ring electrode RF voltage tuning stud (with air deflector
removed)
Tuning
Stud