![Thermo Finnigan LCQ Deca XP Plus Скачать руководство пользователя страница 149](http://html.mh-extra.com/html/thermo-finnigan/lcq-deca-xp-plus/lcq-deca-xp-plus_hardware-manual_1104316149.webp)
Service Maintenance
______________________________________ Replacing PCBs and Assemblies in the Tower
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-43
2. Open the front doors of the MS detector.
3. With a Phillips screwdriver, loosen the two fasteners that hold the front
cover to the embedded computer. Remove the front cover of the
embedded computer.
4. Locate the PCB you want to replace. See Figure 5-14.
5. Disconnect all electrical cables to the PCB that you want to replace.
6. With a Phillips screwdriver, remove the screw that holds the PCB to the
card cage.
CAUTION. Make sure that the XP Plus electronic service switch is in the
Service Position (or shut down the system and disconnect the power cord)
before you proceed.