5. Operating the INFINITE M1000 PRO
2011-09
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
75
Wash
1. Select one of the injectors (Injector A or Injector B) or Injector A and B
2. Select the ‘Piston Strokes’ (1-60).
3. Select the ‘Wash Speed’ (100-300 µl/sec).
4. Select the ‘Refill Speed’ (100-300 µl/sec) or select ‘Refill Speed equal to
Wash Speed’ check box.
5. Press ‘Start Wash’ to start the wash procedure.
6. Click the ‘Save as default’ button to save the selected settings to the
corresponding hardware button (A or B) on the injector box. When using the
hardware buttons for washing these settings will be applied. Start the wash
procedure by pressing and holding the Prime/Wash button for less than 3
seconds.
7. Select ‘Close’ to exit the dialog box.
STOP
Caution
The injector carrier must be in the service position
for the action ‘wash’.
Do not perform wash when the injector is in the instrument.
STOP
Important
Be sure to run a final wash procedure with distilled water and empty the
injector system. For good care and lifetime fill the injector system with
liquid (water) before turning off the instrument.
STOP
Important
Take good care of the injectors, because if they are damaged the
accuracy of dispensing may be affected. This can result in damage to
the instrument.
Note:
Injector needles can be replaced by exchanging the injector together
with the corresponding tubing.
STOP
Important
The button(s) on the injector box include two functions:
Press the button for more than 3 seconds to start ‘prime’.
Press the button for less than 3 seconds to start ‘wash’.
The parameters have to be set in the i-control software.