6. Instrument Specifications
90
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
2011-09
6.3.3
Fluorescein (Fluorescence Intensity) Bottom
Plate type (number of wells)
384
1536
Dispensed Volume [µl]
100
10
Flashes per Well
12
12
Flash frequency
100 Hz
100 Hz
Fluorescein Detection Limit [pM]
≤
35
≤
40
6.3.4
Europium (Time Resolved Fluorescence)
Plate type (number of wells)
384
1536
Dispensed Volume [µl]
100
10
Flashes per Well
12
12
Flash frequency
100 Hz
100 Hz
Europium Detection Limit [pM]
≤
0.75
≤
0.90
6.3.5 HTRF
®
(Time Resolved Fluorescence)
Plate type (number of wells), white
384
Dispensed Volume [µl]
50
Integration Time [µs]
500
Lag Time [µs]
60
Flashes per Well
50
Flash frequency
100 Hz
Signal/Background-Ratio
≥
4
Delta F, Low Calibrator [%]
≥
15
Delta F, High Calibrator [%]
≥
600