8. Cleaning & Maintenance
124
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
2011-09
8.3 Instrument
Decontamination/Disinfection
WARNING
ALL PARTS OF THE INSTRUMENT THAT COME INTO CONTACT WITH
POTENTIALLY INFECTIOUS MATERIAL, BIOLOGICAL SAMPLES, PATIENT
SAMPLES, POSITIVE CONTROL SAMPLES OR ANY HAZARDOUS MATERIAL
MUST BE TREATED AS POTENTIALLY INFECTIOUS AREAS.
IT IS ADVISABLE TO ADHERE TO APPLICABLE SAFETY PRECAUTIONS,
(INCLUDING THE WEARING OF POWDER-FREE GLOVES, SAFETY
GLASSES, SURGICAL MASK AND PROTECTIVE CLOTHING) TO AVOID
POTENTIAL INFECTIOUS DISEASE CONTAMINATION WHEN PERFORMING
THE DECONTAMINATION/DISINFECTION PROCEDURE.
Warning
It is very important that the instrument is thoroughly
decontaminated/disinfected before it is removed from the laboratory or
before any service is performed on it.
STOP
Caution
Ensure that the microplate is removed from the instrument before it is
prepared for shipment. If a microplate is left in the instrument, fluorescent
solutions may spill onto the optical parts and damage the instrument.
Before the instrument is returned to the distributor or service center, all outer
surfaces and the plate transport must be decontaminated/disinfected and a safety
certificate must be completed by the operating authority. If a safety certificate is
not supplied, the instrument may not be accepted by the distributor or service
center or custom authorities may hold it.
8.3.1 Decontamination/Disinfection
Solutions
The instrument (Front, Cover, Plate transport) should be
decontaminated/disinfected using one of the following solutions:
•
Areades
B (Schülke & Mayr GmbH, A-1070 Wien)
•
Dodacarna rapid neu (Schülke & Mayr GmbH, A-1070 Wien)
STOP
Caution
The Decontamination/Disinfection procedure should be
performed by authorized trained personnel in a well ventilated
room wearing powder-free gloves, safety glasses, surgical mask
and protective clothing.
Please note that the decontamination/disinfectant solution can
influence the performance of your instrument if applied inside the
instrument.