5. Operating the INFINITE M1000 PRO
2011-09
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
81
5.7.6
Injector Cleaning and Maintenance
The required maintenance may vary with your application. The following
procedures are recommended for optimal performance and maximum life of the
injector system.
Daily Maintenance:
If not otherwise stated by the manufacturer of the kit to be used, at least the
following tasks must be performed daily:
•
Inspect the pump(s) and tubing for leaks.
•
Flush the whole system thoroughly with distilled or deionized water after
each use and when the pump is not in use. Failure to do so can result in
crystallization of reagents. These crystals can damage the syringe seal and
valve plug resulting in leakage.
STOP
Caution
Do not allow the pump(s) to run dry for more than a few cycles.