6. Instrument Specifications
2011-09
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
93
6.6.3
Dual-Color Luminescence (e.g. BRET)
STOP
Caution
Switch on the instrument at least 15 minutes before starting a
luminescence measurement to ensure stable conditions for the
measurement.
Parameters Characteristics
Integration Time:
100 - 20000 ms
Different integration times are possible for each
wavelength.
Plate Type:
96, 384, and 1536-well microplates
Dynamic Range:
6 decades
6.7 AlphaScreen/AlphaLISA
STOP
Caution
Switch on the instrument at least 15 minutes before starting a
AlphaScreen/AlphaLISA measurement to ensure stable conditions for the
measurement.
Parameters Characteristics
Excitation Time:
Integration Time:
10 - 1000 ms
100 - 20000 ms
Plate Type:
96, 384, and 1536-well microplates
(white or light gray)
Omnibeads Detection Limit
[ng/ml] (384-well)
≤
12.5
Uniformity [% CV]
(384-well)
≤
8
Crosstalk [%]
(384-well)
≤
0.2