8. Cleaning & Maintenance
2011-09
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
123
8. Cleaning
&
Maintenance
8.1 Introduction
STOP
Caution
Ensure that the microplate is removed from the instrument before it is
prepared for shipment. If a microplate is left in the instrument,
fluorescent solutions may spill onto the optical parts and damage the
instrument.
The cleaning and maintenance procedures are important in order to prolong the
instrument's life and to reduce the need for servicing.
STOP
Caution
Keep the plate transport clean, especially take care of the clip
mechanism to fix microplates. Insufficient plate fixing lead to instrument
damages. Strong soiling needs servicing.
This section contains the following procedures:
•
Liquid
Spills
•
Instrument
Decontamination/Disinfection
•
Decontamination/Disinfection
Procedure
•
Safety
Certificate
•
Disposal
8.2 Liquid
Spills
1. Wipe up the spill immediately with absorbent material.
2. Dispose of contaminated material appropriately.
3. Clean the instrument surfaces with a mild detergent.
4. For biohazardous spills, clean with a 5-10 % solution of bleach in
de-ionized water.
5. Wipe cleaned areas dry.