5. Operating the INFINITE M1000 PRO
80
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
2011-09
Using the Injection Strip:
Figure 41
Plate:
Select an appropriate plate type.
Part of the plate:
Select the wells to be dispensed
Well:
The well strip is mandatory. Injection is only possible with a ‘well’ strip.
Injection:
Set up the injection parameters.
Only one injector can be selected per strip. If both injectors are required an
additional injection strip has to be inserted.
Injection volume:
The dispense volume depends on the microplate type. The
plate definition files include the working volume of the microplate. This working
volume defines the maximum volume to be dispensed into the selected
microplate. Therefore, ensure that the selected plate definition file contains the
correct value for the working volume. The maximum dispense volume is also
limited by the installed syringe size (500 µl; 1 ml; 2.5 ml).
Measurement strip (Absorbance in the presented example):
It is mandatory
to use at least one measurement strip in combination with the injection strip. The
position of the measurement strip(s) (before and/or after the injection strip)
depends on the application.
Note
Ensure that the correct value for the Working Volume is inserted in the
plate definition file.