PM (preventive maintenance) reminder
Interval: 1 to 24 months in 1 month increments
Enable: On or off
Custom keypad code
001 to 899 in increments of 1
Custom clinician code
001 to 899 in increments of 1
Custom administrator code
001 to 899 in increments of 1
Date format
US standard (mm/dd/yy)
European standard (dd/mm/yy)
Japanese standard (yy/mm/dd)
Time format
00:00 to 23:59 military or 12-hour am/pm
Downstream occlusion sensitivity
High Sensitivity:
When the high pressure alarm threshold is reached, the downstream occlusion
alarm is triggered immediately.
Low Sensitivity:
When the high pressure alarm threshold is reached, the downstream occlusion
alarm is delayed for 2 seconds. This allows for the pressure to stabilize before a possible alarm. If the
pressure stabilizes below the high pressure alarm threshold before the 2 second delay is complete, the
alarm will not occur.
Upstream occlusion sensor
On or off
NOTE:
The upstream occlusion sensor is automatically disabled during use with medication cassette
reservoirs.
PCA Delivery Specifications
Programming units
If programming through manual mode. Otherwise the programming units are preset through the
CADD
®
-Solis Medication Safety Software.
•
Milliliters (mL)
•
Milligrams (mg)
•
Micrograms (mcg)
Concentration
mg/mL:
•
0.1 to 0.5 mg/mL in increments of 0.1 mg/mL
•
0.5 to 1 mg/mL in increments of 0.5 mg/mL
•
1 to 15 mg/mL in increments of 1 mg/mL
•
15 to 100 mg/mL in increments of 5 mg/mL
mcg/mL:
•
1 to 15 mcg/mL in increments of 1 mcg/mL
•
15 to 100 mcg/mL in increments of 5 mcg/mL
•
100 to 500 mcg/mL in increments of 100 mcg/mL
Continuous rate
0 to 100 mL/hr (or the mg or mcg equivalent)