Alarm Settings
You may adjust various alarm settings through the Alarm Settings task in the Advanced Tasks menu.
Some of the adjustable settings include whether or not an alarm is active, the type of alarm that sounds,
and at what point the alarm sounds. See
on page 110 for more information.
There are 4 types of alarms:
•
High Priority
•
Medium Priority
•
Low Priority
•
Informational
In the Advanced Tasks menu, press
U
or
D
to highlight
Alarm
Settings
and press
S
.
Infusion Alerts On/Off
This alarm setting is available for Step delivery mode only. Turning this setting On results in infusion
notification screens 5 minutes before each automatic step up occurs.
To set the infusion alerts:
1. Make certain the pump is stopped.
2. In the Alarm Settings menu, press
U
or
D
to highlight
Infusion
Alerts On/Off
and press
S
.
3. Unlock the keypad.
4. Press
U
or
D
to highlight
Off
or
On
and select
Save.
Pump Stopped Alarm Type
The pump stopped alarm allows you to select either a high priority or informational alarm as a reminder
that the pump is stopped.
To set the pump stopped alarm type:
1. Make certain the pump is stopped.
2. In the Alarm Settings menu, press
U
or
D
to highlight
Pump
Stopped Alarm Type
and press
S
.
3. Unlock the keypad.
4. Press
U
or
D
to highlight
Informational
or
High Priority,
and
select
Save.