Power Up
Press and hold the power switch to turn the pump on. The pump starts the power up sequence during
which it performs various self-tests, tests for alarm conditions. Watch for the following during power-up:
•
Both the green and amber indicator lights flash.
•
The display quickly flashes gray, then blue. An amber swirl then fills the display, followed by a CADD
®
-
Solis VIP Ambulatory Infusion System display. Look for any stripes or black or white pixels which
indicate a faulty display. If you see any indication of a faulty display, remove the pump from service and
contact Smiths Medical Customer Service.
•
After the power up is completed, listen for the Morse Code “OK” sound (a series of six audible beeps).
If you do not hear this sound, there may be a problem with the audible alarms. If you believe there is a
problem, remove the pump from service and contact Smiths Medical Customer Service.
•
If any issues are found while the pump is performing the self tests, alarms will sound (for example, if
the battery is low or a key on the keypad is stuck in the pressed position).
CAUTION:
If the power up results in an error message indicating that the protocol library was lost,
do not proceed with using the pump. Follow your facility’s procedures for downloading protocol
libraries.
Power Down
Press the power switch to turn the pump off. You must acknowledge the “Power down?” prompt by
selecting
Yes.
The pump then powers down. Be aware of the following conditions when the pump is
powered down:
•
The display is blank.
•
Keypad presses are not detected.
•
The remote dose cord does not function.
•
Medication is not infused.
•
Alarm conditions are not sensed.
•
Audio and visual alarms are not displayed.
•
Communication with CADD®-Solis Medication Safety Software is not possible.
•
If the AC adapter is plugged in, the blue AC power light is on. No other lights are lit.
•
The internal battery used to retain time and date does not charge.
•
If the pump is connected to AC power and a rechargeable battery pack is installed, the battery pack
continues to charge.