Cleaning the Pump and Accessories
NOTE:
Refer to the instructions for use for each accessory before proceeding with cleaning.
The following solutions may be used to clean the pump and accessories, unless otherwise specified:
•
Soap solution
•
Benzalkonium chloride concentrate (0.13%)
•
Glutaral concentrate, USP (2%)
•
10% solution of household bleach (1 part household bleach to 9 parts water)
•
Alcohol, USP (93%)
•
Isopropyl alcohol, USP (99%)
•
Chlorhexidine gluconate (4%)
•
PDI
®
Super Sani-Cloth
®
•
Madacide (MADA Medical)
•
Virex
®
II (Johnson Wax)
•
Coverage
®
Spray and Coverage
®
HB Plus (Steris)
•
CaviCide
®
(Metrex)
•
Quik Fill Compac A-456-N (Airkem)
CAUTION:
•
Do not immerse the pump in cleaning fluid or water. Do not allow solution to soak in to the pump,
accumulate on the keypad, or enter the battery compartment, USB port, remote dose cord jack, or
power jack areas. Moisture buildup inside the pump may damage the pump.
•
Do not clean the pump with acetone, other plastic solvents, or abrasive cleaners, as damage to the
pump may occur.
1. Ensure that debris is not allowed to build up on the pressure plate surface of the pumping mechanism.
Consult the Technical Manual for a diagram of the pumping mechanism. Inspect the air detector
sensor slot and remove any debris.
2. Dampen a soft, lint-free cloth with cleaning solution. Apply the solution to the exterior surface of the
pump or accessory according to manufacturer’s instructions.
Do not allow the solution to soak into the
pump or accessory.
3. Allow the pump to dry completely before use.