28
Pump Programming
Pu
mp
Pr
og
ra
mmi
ng
8. The “Review pump settings” screen appears. Select
Review.
9. Press
U
or
D
to scroll through and review all the settings. If the
highlighted setting is correct, press
Accept Value.
A green check
mark appears next to the setting.
If the highlighted setting is incorrect, press
S
to adjust the setting.
Press
U
or
D
to modify the setting and then press
Save.
10. When the review is complete, select
Next.
11. When a cassette is attached, latched, and locked (if required by
PCA therapies), the "Prime tubing?" screen
may
appear. For more
information on priming, see
The “Start pump?” screen appears. Select
Yes
to start the pump
immediately or
No
to start the pump at a later time.
Manually Programmed Protocols
NOTE:
Manual programming is for unusual circumstances when an order is received which is different
than any of your standard protocols. Programming manually does not use all of the medication safety
features built into the CADD
®
-Solis pump. If you find that you frequently program manually, consider
contacting your CADD
®
-Solis system administrator to discuss if additional protocols should be
created.
Manually programming a protocol or making changes to an individual protocol for a single use may occur
when the doctor’s orders do not match any of the protocols in the library, or when a protocol library is not
available.
Unlike protocols that are created and downloaded into the pump by the CADD
®
-Solis Medication Safety
Software, manual programming allows you to choose units and concentration (in PCA mode), and does
not contain any programming limits. Many of the mode-independent parameters may remain the same
as the parameters from the previously used protocol. For example, if the previously used protocol had the
pump stopped alarm type set as informational, the manual program also has the pump stopped alarm
type set as informational. However, the actual programming limits of the previous protocol are cleared.
The hard and soft limits are all set to the factory default, which means the delivery ranges are not limited.
See the following chart for details.
NOTE:
If the pump was set to the factory default, or if you are using the pump for the first time, the
previously used protocol is the factory default settings. Refer to the tables beginning on page 140.
The following table lists the initial pump settings when programming manually.
Attributes not listed in
the table remain the same as the previously used protocol.