Pump Programming
Pu
mp
Pr
og
ra
mmi
ng
Delivery Hard and Soft Limits
Pump protocol libraries consisting of therapies, qualifiers, and drugs are created and loaded on the pump
using the CADD
®
-Solis Medication Safety Software. Many of the delivery settings have hard and soft limits
assigned to them by the CADD
®
-Solis system administrator. In the unlikely event that any of these values
need to be changed manually on the pump, it can be done under the Delivery Hard and Soft Limits setting
by a user with the administrator code.
The delivery hard and soft limits define the ranges that are available when programming delivery settings
in the View Delivery Settings menu. Not all of the settings have hard and soft maximum and minimum
values. However, the settings still effect the limits that appear in the View Delivery Settings menu. For
example, in PCA mode, the Delivery Limit Period, Delivery Limit Amount, and Max Doses/Hr may not
appear, based on the value of the Delivery Limit Method.
Default delivery hard and soft limits are listed in the table
Manual Mode Initial Delivery Hard and Soft
on page 142. The settings are defined below:
•
Hard Maximum:
The maximum value for the parameter being edited. It is not possible to program the
setting above the hard maximum value.
•
Soft Maximum:
The soft maximum value should be the highest common value for the setting you are
programming. The setting may be programmed above this value, although extra confirmation is required.
•
Current Value:
Displays the current value for the setting in the therapy. Changing this setting changes
the programmed value in the current therapy.
•
Soft Minimum:
The soft minimum value should be the lowest common value for the setting you are
programming. The setting may be programmed below this value, although extra confirmation is required.
•
Hard Minimum:
The minimum value for the parameter being edited. It is not possible to program the
setting below the hard minimum value.
To set the delivery hard and soft limits:
1. Make sure the pump is stopped.
2. In the Advanced Tasks menu (see page 99), press
U
or
D
to highlight
Delivery Hard and Soft Limits
and press
S
.
3. Press
U
or
D
to scroll to and highlight the setting you want to
adjust and press
S
. In this example, only 1 setting is adjustable.
4. Unlock the keypad.
5. Press
U
or
D
to scroll to and highlight the setting you want to
adjust and press
S
.
6. Press
U
or
D
to increase or decrease the limit for the setting and
select
Save.
7. Continue to scroll through the settings until you have set limits for all
the values that you want to set. Select
Back
to go back to the Delivery
Settings Limits menu. Press
U
or
D
to scroll to any other setting
you want to adjust, or press
Back
to finish.