Taper Delivery Mode
Ta
pe
r
Pr
og
ra
mmi
ng
Stopping and Restarting the Pump During an Infusion
If you stop the pump before a taper infusion was fully delivered, you may choose to either restart the
infusion from where delivery left off, or you may reset the infusion.
When you press stop/start
9
, a screen appears with the message, “Infusion was interrupted before
completion. Finish interrupted infusion or start from beginning with new bag.” Select
New Bag
to reset
the reservoir volume and reset the infusion. Or, select
Finish
to resume the infusion from the point it was
at when it was last running.
Taper Down Now
This task allows you to stop an infusion early by immediately tapering down. The following conditions are
necessary for Taper Down Now:
•
The pump must be running.
•
The pump must be delivering at the plateau rate or already be in taper down.
To access the Taper Down Now option:
1. From the Tasks menu (see page 88), press
U
or
D
until
Taper Down Now
is highlighted, and then
press
S
.
NOTE:
If a taper down was not
programmed, you can edit the setting by
pressing
S
. Enter the security code,
and then press
U
or
D
to set the
taper down duration. Press
Confirm
to
accept the change.
2. Select
Yes
to continue with taper down now or press
S
to edit the
taper down period. Editing the taper down now period only affects the
period for this immediate taper. It has not effect on the taper period of
the protocol that is seen in the delivery settings.
NOTE:
Once you select
Yes,
you can
not
restart delivery at the plateau
rate without resetting the infusion period.
3. When “Beginning Taper Down…” appears, select
OK.