Intermittent Delivery Mode
In
te
rmi
tte
nt
Pr
og
ra
mmi
ng
Intermittent Delivery Mode
Intermittent delivery allows the infusion of a specific volume of drug at a regular, programmed interval
and over a specified duration. A dose may be repeated in a cycle up to 96 hours. The KVO (keep vein open)
feature allows delivery of a minimal amount of drug between doses to maintain catheter patency. You may
also delay the start of the first or next dose by using the Next Dose Start Time task.
The following graph illustrates an example intermittent delivery:
Home Screen
The delivery mode and status of the infusion are indicated on the home screen with messages appearing
in text, a graphic in a shape unique to the therapy, and with the colors green, red, and gray. The
colors indicate that the pump is running (green), or stopped (red). If a new patient or new infusion is
programmed but
the pump has not been started,
or if a next dose start time was programmed and the
next dose start time passed while the pump was stopped and not yet started, the graphic is entirely gray.
The location of the color on the graphic also indicates the status of the infusion. The examples below are
specific to the intermittent delivery mode.
•
The pump is running with a manually programmed next dose start time.
•
The pump is running and delivering a dose.
•
The pump is stopped between doses while at the KVO rate.
Intermittent delivery mode graphics do
not
depict the following:
•
The relative ratio of dose period to dose cycle.
•
The number of doses remaining in the current reservoir volume.
Time
Deliv
er
y R
at
e
Delayed start
(optional)
KVO
KVO
Dose
Dose
Duration
Duration
Cycle