Step Delivery Mode
St
ep
Pr
og
ra
mmi
ng
Starting Each infusion
When a new cassette is attached to the pump at the beginning of infusion, the reservoir volume should
be reset.
This also resets the infusion.
Follow the instructions in
beginning on page 78,
for attaching a cassette and resetting the reservoir. When the pump is started, delivery will start at the
beginning of the infusion.
Stopping and Restarting the Pump During an Infusion
If you stop the pump before a step infusion was fully delivered, you may choose to either restart the
infusion from where delivery left off, or you may reset the infusion.
When you press stop/start
9
, a screen appears with the message, “Infusion was interrupted before
completion. Finish interrupted infusion or start from beginning with new bag.” Select
New Bag
to reset
the reservoir volume and reset the infusion. Or, select
Finish
to resume the infusion from the point it was
at when it was last running.
Step Up
This option allows you to increase the rate of infusion at a time other than what is programmed. You
cannot step up to a higher rate than the plateau rate.
To step up from the current rate:
1. In the Advanced Tasks menu, press
U
or
D
to highlight
Step Up
and press
S
.
2. Unlock the keypad.
3. The pump displays a screen that tells you the current rate, and asks you
if you want to step up to the next step rate. Select
Yes.
Time
Deliv
er
y R
at
e
Plateau rate
Initial rate
Rate (step)
increase
Rate (step)
increase
KVO
Infusion volume
Step
duration
New cassette attached.
Reservoir volume and infusion profile reset.