Troubleshooting
Issue
Possible Solution
A continuous two-tone
alarm is sounding, and the
amber light is lit or flashing.
Delivery has stopped. Read the message on the display and refer
to the list of messages in the table below. If the display is blank or
contains random characters, the 4 AA batteries or the rechargeable
battery pack may be depleted. Install 4 new AA batteries or a
rechargeable battery pack.
The pump is sounding
2 beeps every 2 seconds, and
the amber light is flashing.
Look at the message on the display and refer to the list of messages in
the table below.
Three beeps sound every
5 minutes.
This is a reminder that the pump is stopped.
After installing 4 new AA
batteries and powering up
the pump, no screen appears
and no beep sounds.
The batteries may be installed incorrectly. Review the procedure for
installing batteries. Be sure to match the polarity (+ and –) markings
inside the battery door with the markings on the batteries. If there is
still no power, the batteries may be completely depleted. Install 4 new
AA batteries.
Alarms and Messages, Alphabetical List
Alarm / Message
Alarm
Priority
Description / Corrective Action
(Screen is blank and
alarm is sounding)
High
The pump was delivering and the batteries were removed or the
battery door was opened. The pump lost power and is no longer
delivering. Clear the alarm by turning the pump back on, or the
alarm will stop after the power has been off for a minimum of
2 minutes.
A setting was edited,
but not saved, and
the edit was lost.
Medium A parameter was being manually edited, but it was not saved and
the pump reverted to the home screen. Select
Acknowledge
to
clear the alarm and if appropriate, edit the parameter and save.
AC adapter
disconnected.
Low
The AC adapter was disconnected and the pump is being powered
by the 4 AA batteries or the rechargeable battery pack. Select
Acknowledge
to clear the alarm, or the alarm will automatically
clear after 5 seconds. If desired, reconnect the AC adapter.
Air in-line detected.
Press ‘Acknowledge’
then prime tubing.
High
The air detector has detected air in the fluid path. The fluid path
may contain air bubbles. The pump was delivering and is now
stopped and will not run. Select
Acknowledge
to clear the alarm.
If the fluid path contains air bubbles, close the clamps and
disconnect the fluid path from the patient. Follow the instructions
for removing air by priming (see page 83 for more information on
priming). Restart the pump.
Battery depleted.
Pump stopped.
High
The pump was delivering but is now stopped and the battery
power is too low to operate the pump. If the AC adapter is
attached, select
Acknowledge
to clear the alarm.
Remove the batteries. Install 4 new AA batteries or a rechargeable
battery pack. To start delivery, good batteries must always be
installed even when an external source of power is connected. If
appropriate, restart the pump.
Battery depleted.
Pump will not run.
Medium The battery power is too low to operate the pump. If the AC
adapter is attached, select
Acknowledge
to clear the alarm.
Remove the batteries. Install 4 new AA batteries or a rechargeable
battery pack. To start delivery, good batteries must always be
installed even when an external source of power is connected. If
appropriate, restart the pump.