Given and PCA Dose Counters
NOTE:
This report applies only to PCA therapies. See
on page 95 for reports specific to therapies
other than PCA.
This report shows the total amount of drug given and the number of PCA doses given and attempted since
the date and time indicated, which is the last time the information was cleared manually, or when a new
protocol or new patient was started.
•
Total Given
shows the amount of drug (in programmed units) delivered by continuous rate,
clinician boluses, and PCA doses.
•
PCA Doses Given
shows the number of PCA doses actually delivered to the patient, including any
doses stopped in progress.
•
PCA Doses Attempted
shows the total number of PCA doses attempted by the patient while the
pump was running, including those that were delivered, locked out, and stopped in progress.
To clear total given and/or the dose counters:
1. In the View Reports menu, press
U
or
D
to highlight
Given and
PCA Dose Counters
and press
S
.
2. Press
U
or
D
until the counter you wish to clear is highlighted.
3. Select
Clear Given or Clear Doses
to
clear the counter. This also updates the
time and date on this screen.
PCA Dose Graph
NOTE:
This report applies only to PCA therapies. For all other therapies, information is available in the
report
Delivery History and Pie Chart.
This graph displays the number of doses attempted and given in 30 minute increments, starting from the
current time to 8 hours in the past, or since the start of a new protocol or new patient. Select this report to
review the number of attempted doses for a particular time frame.
To view the PCA Dose Graph:
In the View Reports menu, press
U
or
D
to highlight
PCA Dose Graph
and
press
S
.
In this example, the patient has attempted 3
PCA doses, and 1 was delivered.